Electrical Connections - E+E Elektronik EE22 Series Manual

Humidity / temperature transmitter
Table of Contents

Advertisement

Available languages

Available languages

4.
EE22-xFT1,2,3x
+
power supply
~
for 0 - 1 V:
10 - 35V DC
9 - 29V AC
-
for 0 - 5 V:
12 - 35V DC
~
15 - 29V AC
for 0 - 10 V:
15 - 35V DC
15 - 29V AC
EE22 with connectors
EE22-xFT3x
Plug for supply and
analogue output
(front view)
EE22-xFT6x
Plug for supply and
analogue output
(front view)
EMC - Recommendations for Wiring
The indicated EMC guidelines can be guaranteed only in compliance with the probe
cables which are available as accessories.
Metal housing models (EE22-MFTx) must be grounded for safe operation either inside
at the designated points or outside at the mounting plates.
The probe has to be mounted electrically isolated. --> No grounding!
Seperate cables which generate electromagnetic disturbances from EE22, sensing
probe cable and analogue output cabels.
Shields of analogue signal cables should be grounded on one end only - best to the
connected signal processing device. Unshielded cables should be twisted together in order
to reduce symmetrical electromagnetic disturbances.
Keep all cables as short as possible. Loops of cables may reduce EMC performance.
Ground unused wires at both ends. Install all cables as close as possible to
ground potential (e.g.: close to the walls, mounting plates or steel structure elements).

ELECTRICAL CONNECTIONS

V+
1
GND
2
RH
3
T
4
V
V
EE22-xFT6x
power supply
10...35V DC
Uv - 10V
[W]
R <
L
0.02 A
Description:
GND
V+
T
Euro Standard
Description:
not assigned
V+
T
Euro Standard
14
+
-
mA
mA
Description:
4
3
not assigned
5
2
1
RH
Description:
4
3
not assigned
5
1
2
RH
1
V+
2
RH
3
T

Advertisement

Table of Contents
loading

Table of Contents