E-Series E9300 Average Power Sensor Specifications - Keysight N1914A Datasheet

Power meters
Hide thumbs Also See for N1914A:
Table of Contents

Advertisement

E-Series E9300 Average Power Sensor Specifications

The E-Series E9300 wide dynamic range, average power sensors are designed for use with the EPM family of power meters� These
specifications are valid ONLY after proper calibration of the power meter and apply for CW signals unless otherwise stated�
Specifications apply over the temperature range 0 to 55 °C unless otherwise stated, and specifications quoted over the temperature
range 25 °C ± 10 °C, conform to the standard environmental test conditions as defined in TIA/EIA/IS-97-A and TIA/EIA/IS-98-A�
The E-Series E9300 power sensors have two independent measurement paths (high and low power paths) as shown in Table 7�
Table 7. E9300 Series two-path specification.
High power path
Low power path
Table 8. E9300 Series sensors specification.
Model
Frequency
range
–60 to +20 dBm wide dynamic range sensors
E9300A
10 MHz to 18 GHz
E9301A
10 MHz to 6 GHz
E9304A
9 kHz to 6 GHz
–30 to +44 dBm wide dynamic range sensors
E9300B
10 MHz to 18 GHz
E9301B
10 MHz to 6 GHz
–50 to +30 dBm wide dynamic range sensors
E9300H
10 MHz to 18 GHz
E9301H
10 MHz to 6 GHz
Find us at www.keysight.com
"A" suffix sensors
–10 to +20 dBm
–60 to –10 dBm
Maximum SWR
Maximum SWR
(25 °C ± 10 °C)
(0 to 55 °C)
10 to < 30 MHz : 1.15
10 to < 30 MHz : 1.21
30 MHz to < 2 GHz : 1.13
30 MHz to < 2 GHz : 1.15
2 to < 14 GHz : 1.19
2 to < 14 GHz : 1.20
14 to < 16 GHz : 1.22
14 to < 16 GHz : 1.23
16 to 18 GHz : 1.26
16 to 18 GHz : 1.27
10 to < 30 MHz : 1.15
10 to < 30 MHz : 1.21
30 MHz to < 2 GHz : 1.13
30 MHz to < 2 GHz : 1.15
2 to 6 GHz : 1.19
2 to 6 GHz : 1.20
9 KHz to < 2 GHz : 1.13
9 KHz to < 2 GHz : 1.15
2 to 6 GHz : 1.19
2 to 6 GHz : 1.20
10 MHz to < 8 GHz : 1.12
10 MHz to < 8 GHz : 1.14
8 to < 12.4 GHz : 1.17
8 to < 12.4 GHz : 1.18
12.4 to 18 GHz : 1.24
12.4 to 18 GHz : 1.25
10 MHz to 6 GHz : 1.12
10 MHz to 6 GHz : 1.14
10 MHz to < 8 GHz : 1.15
10 MHz to < 8 GHz : 1.17
8 to < 12.4 GHz : 1.25
8 to < 12.4 GHz : 1.26
12.4 to 18GHz : 1.28
12.4 to 18GHz : 1.29
10 MHz to < 6 GHz : 1.15
10 MHz to < 6 GHz : 1.17
"B" suffix sensors
+20 to +44 dBm
–30 to +20 dBm
Maximum power
+25 dBm (320 mW) average
+33 dBm peak (2 W)
(< 10 μsec)
+25 dBm (320 mW) average
+33 dBm peak (2 W)
(< 10 μsec)
+25 dBm (320 mW) average
+33 dBm peak (2 W)
(< 10 μsec)
0 to 35 °C: 30 W avg
35 to 55 °C: 25 W avg
< 6 GHz: 500 W pk
> 6 GHz: 125 W pk
500 W.μS per pulse
0 to 35 °C: 30 W avg
35 to 55 °C: 25 W avg
< 6 GHz: 500 W pk
> 6 GHz: 125 W pk
500 W.μS per pulse
3.16 W avg
100 W pk
100 W.μS per pulse
3.16 W avg
100 W pk
100 W.μS per pulse
"H" suffix sensors
0 to +30 dBm
–50 to 0 dBm
Connector type
Type-N (m)
Type-N (m)
Type-N (m)
Type-N (m)
Type-N (m)
Type-N (m)
Type-N (m)
Page 16

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1913a

Table of Contents