N1913A/14A EPM Series Power Meters Performance Characteristics (Continued)
Line power
Input voltage range
Input frequency range
Power requirement
Environmental characteristics
Electromagnetic compatibility
Product safety
Low Voltage Directive
Operating environment
Temperature
Maximum humidity
Maximum altitude
Storage conditions
Non-operating storage temperature
Non-operating maximum humidity
Non-operating maximum altitude
Remote programming
Interface
Command language
GPIB compatibility
1� Characteristics describe product performance that is useful in the application of the product, but is not covered by the product warranty�
Find us at www.keysight.com
90 to 264 VAC, automatic selection
47 to 63 Hz and 400 Hz at 110 Vac
75 VA (50 Watts)
Complies with the essential requirements of EMC Directive (2004/108/EC) as follows:
IEC61326- 1:2005 / EN61326- 1:2006
CISPR11:2003 / EN55011:2007 (Group 1, Class A)
The product also meets the following EMC standards:
Canada: ICES/NMB- 001:2004
Australia/New Zealand: AS/NZS CISPR 11:2004
This product conforms to the requirements of the following safety standards:
IEC 61010- 1:2001 / EN 61010- 1:2001
CAN/CSA- C22.2 No.61010- 1- 04
ANSI/UL61010- 1:2004
This product conforms to the requirements of European Council Directive "2006/95/EC"
0 to 55 °C
95% at 40 °C (non-condensing)
4,600 meters (15,000 feet)
–40 to +70 °C
90% at 65 °C (non-condensing)
4,600 meters (15,000 feet)
GPIB, USB and LAN interfaces operates to IEEE 488.2 standard
SCPI standard interface commands. Code-compatible with legacy E4418B/9B EPM Series, 436A, 437B
and 438A power meters (43X compatibility only with option N191xA-200).
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0