E-Series E9300 Average Power Sensor Specifications (Continued)
Power linearity
1
Table 9. E9300 Series power linearity (after zero and cal at ambient environmental conditions) sensor.
Sensor
E9300A, E9301A, E9304A
E9300B, E9301B
E9300H, E9301H
1� After zero and calibration at ambient environmental conditions�
0.5
0.4
0.3
0.2
0.1
0
–0.1
–0.2
–0.3
–0.4
–0.5
–30
–26
–22
–18
–14
1
0.8
0.6
0.4
0.2
0
–0.2
–0.4
–0.6
–0.8
–1
–10
–5
0
1
0.8
0.6
0.4
0.2
0
–0.2
–0.4
–0.6
–0.8
–1
–10
–5
0
Find us at www.keysight.com
Power
–60 to –10 dBm
–10 to 0 dBm
0 to +20 dBm
–30 to +20 dBm
+20 to +30 dBm
+30 to +44 dBm
–50 to 0 dBm
0 to +10 dBm
+10 to +30 dBm
–10
–8
–2
2
6
10
14
Power (dBm)
5
10
15
20
25
Power (dBm)
5
10
15
20
25
Power (dBm)
Linearity (25 ± 10 °C)
± 3.0%
± 2.5%
± 2.0%
± 3.5%
± 3.0%
± 2.5%
± 4.0%
± 3.5%
± 3.0%
Typical E9300A/01A/04A power linearity at 25 °C, after zero and
calibration, with associated measurement uncertainty�
Power range
–30 to –20 dBm
–20 to –10 dBm
–10 to 0 dBm
0 to +10 dBm
+10 to +20 dBm
18
20
Typical E9300B/01B power linearity at 25 °C, after zero and
calibration, with associated measurement uncertainty�
Power range
–6 to 0 dBm
0 to +10 dBm
+10 to +20 dBm
+20 to +26 dBm
30
Typical E9300H/01H power linearity at 25 °C, after zero and
calibration, with associated measurement uncertainty�
Power range
–26 to –20 dBm
–20 to –10 dBm
–10 to 0 dBm
0 to +10 dBm
+10 to +20 dBm
+20 to +26 dBm
30
Linearity (0 to 55 °C)
± 3.5%
± 3.0%
± 2.5%
± 4.0%
± 3.5%
± 3.0%
± 5.0%
± 4.0%
± 3.5%
Measurement uncertainty
± 0.9%
± 0.8%
± 0.65%
± 0.55%
± 0.45%
Measurement uncertainty
± 0.65%
± 0.55%
± 0.45%
± 0.31%
Measurement uncertainty
± 0.9%
± 0.8%
± 0.65%
± 0.55%
± 0.45%
± 0.31%