Erratic Controls Display Or Operation; Power Failure During Sterilizing Cycle - VWR AccuSterilizer 1130 Series Operation Manual

Table top steam sterilizer
Table of Contents

Advertisement

Warning
Reprocess your load in
the event the sterilizing
cycle has been terminated
prematurely. The load
may not be sterile when
the sterilizing cycle has
been terminated prema-
turely.
Warning
Reprocess your load in
the event the sterilizing
cycle has been terminat-
ed prematurely. The load
may not be sterile when
the sterilizing cycle has
been terminated prema-
turely.
Erratic Controls Display or
Operation
Abnormal sterilizer operation or controls display
could be caused by an internal or external
electrical voltage surge. Such a voltage surge
could be cause be any of a number of events: a
momentary surge in the power supply, a nearby
lightening storm, or even a static electrical charge
carried by operating personnel.
Abnormal operation or displays are not common
occurrences and may never happen. No damage to
the equipment will occur if such an event should
happen.
If an abnormal condition should occur, terminate
any processing cycle in process. (Press "Stop.")
Reset the sterilizer by turning the power switch (on
the front right corner of the unit) to OFF. Wait 30
seconds and turn the power switch back on. Select
your desired cycle and proceed with normal
operation.
Power Failure During Sterilizing
Cycle
All Selector/Indicator Panel lights will be off.
Allow the unit to cool. Reset the sterilizer by turning
the power switch (under the front right corner of the
unit) to OFF. Wait 30 seconds and turn the power
switch back on. Drain excess water from the
sterilizer chamber. Then reprocess your load
normally.
S
O
S
PECIAL
PERATING
ITUATIONS
45