Program Summary Chart - VWR AccuSterilizer 1130 Series Operation Manual

Table top steam sterilizer
Table of Contents

Advertisement

D
ESCRIPTION

Program Summary Chart

Program
Liquids
Unwrapped
Wrapped
Packs
Optional
Note
When sterilizing handpieces, refer to the manufacturer's specifications on
handpieces for contra indication of sterilization parameters.
12
Temperature
Time
121°C
15 minutes
(250°F)
135°C
3 minutes
(275°F)
135°C
10 minutes
(275°F)
121°C
30 minutes
(250°F)
User defined
User defined
from 100°C to
from 1 to 99
136°C (212°F
minutes
to 277°F)
Drying Time
Items To Be Sterilized
None
Distilled water, media solutions, other
liquid prepared in closed bottles or
flasks (not greater than 2 liters). Not
for clinical applications.
Default:
Unwrapped instruments; metal glass
15 minutes,
and plastic objects. Heat-resistant
rubber tubing which will not be used in
can be set
surgical applications. Any other items
from 1 to 99
for which 135°C for 3 minutes is
minutes
appropriate.
Default:
Wrapped cloths, bandages, pads, fab-
rics, loosely wrapped individual instru-
15 minutes,
ments, layers of instruments separat-
can be set
ed by fabric, wrapped trays of loose
from 1 to 99
instruments, instruments in bags,
minutes
heat-resistant rubber tubing. Any
other items for which 135°C for 10
minutes is appropriate.
Default:
Packs of towels. Groups of instru-
ments in commercially prepared
15 minutes,
packs. Instruments subject to pro-
can be set
longed storage. Surgical gloves
from 1 to 99
wrapped for sterilization. Any other
minutes
items for which 121°C for 30 minutes
is appropriate, except liquids.
User defined
Items appropriate to the user defined
from 1 to 99
parameters. Dental handpieces
should be run in this cycle at 132°C
minutes
for 10 minutes.