Page 1
Instruction manual Microbiological Air Sampler ® SAS Super ISO USB EU Catalogue Number: 710-2088 710-2087 710-2090 710-2089 VERSION: 2 ISSUED: APRIL 2021...
Page 2
Europe VWR International Srl Via S. Giusto 85 20153 Milano Italy Tel.: + 39 0233 203 11 http://it.vwr.com Country of origin Italy vwr.com | Instruction manual...
LEGAL ADDRESS OF MANUFACTURER FUNCTIONS ACCESSORIES 9-14 WARNING “OPTIONS” FUNCTION TECHNICAL SERVICES 15-24 OPERATIONS WARRANTY SAFETY INFORMATION REFERENCES COLONY FORMING UNIT COMPLIANCE WITH LOCAL LAWS CORRECTION FACTOR AND REGULATIONS TECHNICAL FEATURES TROUBLESHOOTING EQUIPMENT DISPOSAL INTENDED USE FIRMWARE UPDATE INSTALLATION Instruction manual | vwr.com...
Warning! This document is the property of VWR International S.r.l. - Milan - Italy It may not be duplicated or distributed without the owner’s authorisation. Patent Pending Manufactured by VWR International S.r.l. - Milan, Italy Safety Information Please use this device only for the purposes indicated.
To organise sequential sampling to obtain a more representative To aspirate air in a laminar flow pattern with sufficient velocity to sample under actual operating conditions. impact organisms on an agar surface. Instruction manual | vwr.com...
Page 6
Visual display (Touch the OK button for at least two seconds). Each time the instrument is turned on, a five second visual display presentation will appear displaying the following information. This information appears in automatic succession. vwr.com | Instruction manual...
(1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60 minutes) OPTIONS To change general settings of the device. See description below BATTERY Press to see the battery status Instruction manual | vwr.com...
Page 8
Press OK to start air sampling with the last cycle. The motor will run until the figures in the brackets reach the same figure as displayed to the right and then it will stop. To stop the run, press and . vwr.com | Instruction manual...
Normal areas (laboratory benches, houses, etc) 200 - 500 litres of air ‒ Sterile or high risk areas (cleanrooms, operating theatres, etc) 500 - 1000 litres of air Press to switch off (from the Main Screen), or each time you need to end an action. Instruction manual | vwr.com...
Page 10
Current aspirated volume of air with progress bar If the following display appears, the Login function is activated. Please use a User ID to login. For further details please go to the LOGIN section in this manual. vwr.com | Instruction manual...
Page 11
(for example 500 litres) Press OK to select the volume and the display will show the selected volume. Press OK to aspirate the selected volume of air. Display will flash every 4 seconds during aspirating of air. Instruction manual | vwr.com...
Page 12
Date and Time Press OK to select the volume and the display will show the selected volume. Press OK to aspirate the selected volume of air. Display will flash every 4 seconds during aspirating of air. vwr.com | Instruction manual...
Page 13
Select OPERATOR function by pressing the “UP” or “DOWN” arrows from a maximum of 20 programmed IDs. Press OK and select OPERATOR by pressing the “UP” or “DOWN” arrows (for example 0000) Press OK to select the OPERATOR, the display will show the selected volume. Instruction manual | vwr.com...
Page 14
Display will flash every 4 seconds during run. “BATTERY” function Select BATTERY function by pressing the “UP” or “DOWN” arrows. Press OK to see the status of the BATTERY, indicated by a status bar. vwr.com | Instruction manual...
20 you can add a new entry, otherwise the ADD? option not shown. You can delete entries as long as there are at least 2, otherwise the DELETE? option is not shown. The maximum air volume is 2000 litres. Instruction manual | vwr.com...
Page 16
20 you can add a new entry, otherwise the ADD? option is not shown. You can delete entries as long as there are at least 2, otherwise the DELETE? option is not shown. vwr.com | Instruction manual...
Page 17
The data is recorded in the file “DISPLAY RECORD”. Select EDIT OPERATORS function by pressing the “UP” or “DOWN” arrows. Press OK to modify the OPERATORS. Instruction manual | vwr.com...
Page 18
Multi-Cycle mode function. An E (Error) will be displayed in the Data if the aspirating Run were aborted by the Operator. Press to exit this function. vwr.com | Instruction manual...
Page 19
Press OK to confirm and store the data on the USB Stick. If the data were successfully saved the following information appears on the display. Press to exit this function. Press to exit this function. Instruction manual | vwr.com...
Page 20
Press OK to modify the date and time. Change the number indicated by “flashing” by pressing the “UP” or “DOWN” arrows and press OK to go to the next number. Repeat until you reach the last number, confirm by pressing OK. vwr.com | Instruction manual...
Page 21
To unlock functions and enable full functionality, select the password (0-9) by pressing the “UP” or “DOWN” arrows. Press the OK button to go to the next digit. Press OK to confirm, if the password is correct, all functions are unlocked. Instruction manual | vwr.com...
Page 22
Operator ID´s can be edited in the options menu (up to 20 operator IDs can be defined). The options menu itself is protected by the ADMIN password. A normal operator cannot see or modify the operator list. Furthermore, the menu to select an operator ID from the operator ID list is removed. vwr.com | Instruction manual...
Page 23
Select “STORE CONFIG” function by pressing the “UP” or “DOWN” arrows and press OK to access this function. Press OK to confirm and store the data on the USB Stick. If the data were successfully saved the following information appears on the display. Press to exit this function. Instruction manual | vwr.com 23...
Page 24
Select “LOAD CONFIG” function by pressing the “UP” or “DOWN” arrows and press OK to access this function. Press OK to confirm and load the config.sas file from the USB Stick. If the data were successfully loaded the information SUCCESS appears on the display. vwr.com | Instruction manual...
X = Colony Forming Units per 1000 litres (= 1 cubic meter) of air X=(80*1000)/200 400 CFU per 1000 litres of air (1000 litres=1m To express the final result in CFU/ft multiply the CFU/m value by 0,02832 Instruction manual | vwr.com 25...
Page 26
55mm contact plates - 90mm Petri Plates - 219 holes 1,00 mm 55mm contact plates - 90mm Petri Plates - 401 holes 0,75 mm 1088 1307 vwr.com | Instruction manual...
Check battery and replace it “LOW BATTERY” message Battery is low Recharge battery “CALIBRATION EXPIRED” message Instrument needs to be calibrated Send the instrument to VWR International or an authorised dealer MOTOR ERROR Motor not working or not connected Connect the motor USB ERROR...
The Firmware in now updated to the latest version and the Air sampler can be used normally. After restart is shown the instrument is switched off and on again. Then you can remove the USK stick vwr.com | Instruction manual...
SAS stainless steel Petri head + adaptor 710-0877 SAS aluminium Petri head + adaptor 710-0879 IQ OQ PQ validation protocols for SAS Super ISO USB 100 and 180 710-0956 Infrared remote control for SAS Super ISO USB 710-0969 Instruction manual | vwr.com 29...
VWR International warrants that this product will be free from defects in material and workmanship for a period of two (2) years from date of delivery. If a defect is present, VWR will, at its discretion and cost, repair, replace, or refund the purchase price of this product to the customer, provided it is returned during the warranty period.
By doing so, you will help to conserve natural and environmental resources and you will ensure that your equipment is recycled in a manner that protects human health. Thank you Instruction manual | vwr.com...
Page 32
Tel.: +47 22 90 00 00 Lutterworth (international) info.no@vwr.com Leicestershire info.fr@vwr.com LE17 4XN * 0,18 € TTC/min + prix appel Tel.: +44 (0) 800 22 33 44 uksales@vwr.com GO TO VWR.COM FOR THE LATEST NEWS, SPECIAL OFFERS AND DETAILS FROM YOUR LOCAL VWR SUPPORT TEAM ZPROVWR-33022-EN...
Need help?
Do you have a question about the avantor 710-2088 and is the answer not in the manual?
Questions and answers