Only VWR's rotors must be used with the unit.
5 Symmetrically load centrifuge tubes into rotor
CAUTION
Make sure the rotor lid is securely fixed on the rotor, as well as the rotor and shaft are
tightened. Otherwise, the rotor may lift off while rotating and cause damage to the
centrifuge and rotor.

Operation

CAUTION
Do not push or lean against the centrifuge while it is running.
Do not run the centrifuge when fragments or sample solutions are left in the centrifuge
chamber. Always keep the centrifugal chamber clean.
If the centrifuge makes strange noise during operation, stop it immediately and contact
our service center. Notify them of the warning code if displayed.
1 Normal operation
Turn on the power switch, centrifuge will display the running interface last time after passing
the self-diagnostic checks, see figure 9-1 below:
Speed: 4500rpm. Running time: 12 minutes.
The door lock is released.
1)
Set the operation programs
Press the
button to select required program. The parameter can be modified when
the program is flashing. Rotate the program button clockwise
value. Rotate the program button clockwise
program button
speed increment is 100 rpm, the minimum time increment is 1 second.
45
×100rpm
Figure 9-1 Last running interface
faster, and the parameter value will increase faster. The minimum
12
min
to increase parameter
to decrease parameter value.Rotate the