Operating Instructions
Removing and Installing the Rotor
Figure 3.1 Removing Shield
1. Remove the shield by lifting it
Figure 3.2 Locking Shield
1. Locked and Unlocked Positions
5
Use an 1/8 inch hex key to remove the hex screw of the rotor hub.
6
Remove the rotor by lifting it straight up and out of the centrifuge.
Reinstalling the Rotor
1
Align the indent on the side of the rotor hub with the flat on the motor shaft.
2
Slide the rotor onto the motor shaft to seat. Do not force the rotor onto the motor shaft.
If it does not slide down easily re-align the rotor with the motor shaft.
3
Reinstall hex screw and tighten with 1/8 hex key (supplied).
4
Reinstall the shield into the housing by grasping the shield in the center and rotating it
counter-clockwise until it is in the locked position.
3-2
straight up and out of the
centrifuge.
55-006276-001CC