Table 2.2
Cycle Time Indicator Combination
Continuously illuminated Error or
Service indicator.
55-006276-001CC
Error Code Table (Continued)
Problem
Latch is not locked.
Rotor exceeded set speed.
Cover is not closed.
Out of balance.
Locked Rotor
Shield is not installed.
Cycle Counter
System Overview
Error Code Table
Solution
Confirm cover is fully seated
in closed position.
Remove and inspect rotor,
reinstall.
Confirm cover is fully seated
in closed position.
Confirm tubes are balanced in
rotor and centrifuge is on a
level surface.
Contact customer service.
Confirm no obstruction in
bowl area.
Confirm shield is installed and
in the locked position.
50,000 spins have been
reached, please conduct
periodic maintenance check.
(For more information, refer
to Maintenance).