Error Messages - Konica Minolta CHROMA METER CS-200 Instruction Manual

Hide thumbs Also See for CHROMA METER CS-200:
Table of Contents

Advertisement

Error Messages

Error message appears on LCD screen for incorrect operation of this instrument through
key. Below table shows type of error message, its description and corrective action
respectively.
Error message
1
BATTERY OUT
2
DATA SET ERROR
3
CH00 DATA
NOT SETTA-
BLE BY USER
4
INCORRECT
OBSERVER
CONDITION
5
NO DATA
6
UNDER
Cause (Description)
Battery voltage decreases.
Entered numerical value is
out of range.
When CH00 is selected,
calibration channel is go-
ing to be operated or target
color is going to be set.
Measurement data measured
with different observer angle
from that for channel is going
to be used for user calibration
for that channel, or for target
color setting.
Without measurement data,
user calibration is going
to be performed, or target
color to be set.
Luminance of measuring
object is lower than avail-
able measurement range.
110
Corrective Action
• Switch this instrument OFF (O
side), and replace with new batter-
ies or use AC adapter.
• Enter again. Note that measure-
ment value should completely
satisfy the ranges below:
0<x<1,
0<y<1,
0<x+y 1,
0<X 99999000000,
0<Y 99999000000,
0 Z 99999000000
• Select channel except CH00 and
redo the operation.
• Select measurement data meas-
ured with channel and observer
angle or reset calibration channel,
and redo setting.
• Select data to measure or meas-
urement data, and redo setting.
• Remove lens cap and check if
luminance of measuring object
is within available measurement
range, and then redo measure-
ment.
• Check if user calibration is cor-
rectly performed, and redo meas-
urement.
• Check if settings for measurement
angle, observer angle and lens are
correct, and redo measurement.
• If symptom does not improve,
please contact the nearest Konica
Minolta-authorized service facility.

Advertisement

Table of Contents
loading

Table of Contents