Error Messages

An error may occur during operation due to some problems. If an error
occurs, refer to the table below and take the necessary action.
Message
Initial communication error No
response from the receptor head
during initial communication
E r r U
Start command error The receptor head
E r r E
has not received the start command,
because, for instance, the receptor head
was not connected to the main body at
the time the power was turned ON.
EEPROM error Error with the EEPROM
E r r
data (in the receptor head)
Preset value error The value is
E r r A
outside the settable range.
Communication error (main body
↔ receptor head) Communication
E r r C
between the main body and receptor
head was not performed properly.
50
Description
Action
Check connection. If the instrument is
connected properly, contact the nearest
Konica Minolta-authorized service facility.
Set the power switch to "I" (ON)
again. If this message reappears,
contact the nearest Konica Minolta-
authorized service facility.
Remove the batteries (or AC adapter),
then re-install them. If this message
reappears, contact the nearest Konica
Minolta-authorized service facility.
Check the settable range, and make
sure that the value is inside the range.
Check connection. If the instrument is
connected properly, contact the nearest
Konica Minolta-authorized service facility.