Trouble Shootings; Error Messages - Konica Minolta JM-103 Instruction Manual

Hide thumbs Also See for JM-103:
Table of Contents

Advertisement

Trouble Shootings

Error Messages

The following warnings may be displayed on the display win-
dow.
Error Message
Cause
T h e m e a s u r e d
value is abnormal.
I n t h e c a s e o f
average measure-
ment, measure-
ment fluctuation is
excessively large.
It is possible that
a m e a s u r e m e n t
error has occurred
d u r i n g a v e r a g e
measurements, or
a malfunction has
arisen in the hard-
ware.
There is a possibil-
ity that charging is
insufficient.
Solution
Place the measuring probe per-
pendicular to the recommended
point (forehead or sternum) and
take measurement again.
Set the POWER switch to OFF,
wait 10 seconds or more, and
then set it to ON again.
If this warning still reappears,
contact the nearest KONICA
MINOLTA SENSING authorized
service facility.
Set the POWER switch to OFF,
and then place the instrument
on the charger unit to charge it.
If this warning still reappears,
contact the nearest KONICA
MINOLTA SENSING authorized
service facility.
29

Advertisement

Table of Contents
loading

Table of Contents