Patient Data Outputs

The Avant 9700 features real-time and print-on-demand capabilities, which can be
accessed from the Parameter Adjustment menu by toggling the Data Output icon
between print-on-demand and real-time modes.
Data Output icon
All outputs include a header containing model number, time, and date information.
The Avant 9700 provides real-time data output capability via the RS232 connector
port. A null modem cable may be used to connect the Avant 9700 to the receiving
computer. The information from the Avant 9700 is sent in an ASCII serial format
at 9600 baud with 8 data bits, 1 start bit, and 2 stop bits. Each line is terminated by
CR/LF.
Print-On-Demand Output (Default)
The Avant 9700 features a Print-On-Demand option, allowing users to output data
each time the ON/STANDBY button is pressed.
• When the Avant 9700 is in Trend Histogram view, all data from the
recording selected on the trend display are output—including the start
time of the record, the record's SpO
rate histogram data, and the record's ending time.
In any other display view, time, date, SpO
addition, space is provided to include patient and physician information.
Real-Time Patient Data Output
Data are sent once per second by the Avant 9700 in one of the following formats:
• If DIP switch 8 is in the UP position, the data will be displayed as follows:
where XXX and YYY are the fast-responding SpO
• If DIP switch 8 is in the DOWN position, the data will be displayed
as follows:
where XXX and YYY are the SpO
DIP switch 2.
Print-On-Demand indicator
and pulse rate data are output. In
2
SPO2=XXX HR=YYY F
SPO2=XXX HR=YYY
and pulse rate values as selected by
2
histogram data, the record's pulse
2
and pulse rate values.
2
Real-time indicator