Serial Patient Data Outputs; Print-On-Demand Output; Real-Time Patient Data Output - Nonin Avant 9700 Operator's Manual

Digital pulse oximeter
Hide thumbs Also See for Avant 9700:
Table of Contents

Advertisement

• Use the Plus (+) and/or Minus (-) buttons to select dEL yES if desired.
• Press the ON/STANDBY button for the next prompt.
• dnE CLr confirms that the memory is clear.
6. Press the ON/STANDBY button to return to normal operation.
7. For more information about using nVISION, refer to nVISION's online help.
NOTES:
• Patient memory cannot be cleared when the Avant 9700 is in Patient Security mode.
• Selecting yES from the dEL window will permanently delete the device's memory.

Serial Patient Data Outputs

The Avant 9700 features real-time and print-on-demand options. All reports include a header
with the model number, time, and date information.
Data Output icon
Real-time data output is provided by connecting a null modem cable to the RS232 connector port
and the receiving computer. Data is sent in an ASCII serial format at 9600 baud with 8 data bits,
1 start bit, and 2 stop bits; each line is terminated by CR/LF.

Print-On-Demand Output

Avant 9700 allows Print-on Demand if the system is connected to a printer. This operation is
performed by pressing the ON/STANDBY button, as needed. The device features a Print-On-
Demand option, allowing users to output data each time the ON/STANDBY button is pressed.
When the Avant 9700 is in Trend Histogram view, all data from the recording selected on the
trend display are output, including the start time of the record, the record's SpO
the record's pulse rate histogram data, and the record's ending time. In any other display view,
time, date, SpO
and pulse rate data are output. In addition, space is provided to include patient
2
and physician information.

Real-Time Patient Data Output

Data output is sent once per second in one of the following formats:
• If option switch 8 is in the UP position, the data will be displayed as follows:
SPO
=XXX HR=YYY F
2
where XXX and YYY are the SpO
• If option switch 8 is in the DOWN position, the data will be displayed as follows:
SPO
=XXX HR=YYY
2
where XXX and YYY are the SpO
NOTES:
• To insert an event marker, press the ON/STANDBY button.
• The serial output rate is automatically returned to its previous setting upon device startup.
26
Print-On-Demand indicator
and pulse rate values and F signifies "output: fast".
2
and pulse rate values as selected by option switch 2.
2
Memory and Data Output Features
Real-time indicator
histogram data,
2

Advertisement

Table of Contents
loading