Option Switches; Table 5. Option Switches - Nonin Avant 9700 Operator's Manual

Digital pulse oximeter
Hide thumbs Also See for Avant 9700:
Table of Contents

Advertisement

Operating the Avant 9700
Function
Enter Patient
Security Mode

Option Switches

The Avant 9700 contains eight option switches located under the back cover (See "Installing the
Batteries" for instructions on removing the back cover). The device is supplied with all option
switches in the DOWN position.
Switch
Switch 1
Alarm Disable
Up—Alarm volume may be disabled
Down—Alarm volume cannot be set to zero (disabled)
Switch 2
Normal / Slow SpO
Up—Slow Averaging (8 beat exponential average)
Down—Normal Averaging (4 beat exponential average)
Switch 3
Alarm Unlocked / Locked
Up—Alarms Locked
Down—Alarms Unlocked
Switch 4
Factory / User-Defined Defaults
Up—User-Defined Defaults for Alarm Limits and Volume Settings
Down—Factory Defaults for Alarm Limits and Volume Settings
Table 4: User-Defined Defaults (Continued)
Button
then
then
+
Table 5: Option Switches
Averaging
2
Instruction
• After entering the Parameter Adjustment
menu via the Settings button, use the
Right Control button to scroll to the
Patient Security Mode icon.
• When the Patient Security Mode icon is
highlighted, press and hold the Plus and
Minus buttons simultaneously for
3 seconds.
• To deactivate Patient Security mode,
highlight the Patient Security Mode icon
and press and hold the Plus and Minus
buttons simultaneously for 3 seconds.
Function
15