Wipe disinfection for the camera and prismatic tube

NOTE
Do not spray the CEREC Bluecam with or immerse it in cleaning
agents or disinfectants!
1. Remove the camera support (if present).
2. Carry out a preliminary cleaning process (see "Pre-cleaning of the
prismatic tube  [→ 22]").
3. Use a new absorbent cotton gauze which is moistened with one of
the agents listed in the "Cleaning agents and disinfectants  [→ 3]"
section. First wipe the camera enclosure and then the prismatic
tube. The camera enclosure and prismatic tube are disinfected
through this step.
4. Wipe down the prismatic tube afterwards again with the absorbent
cotton gauze dipped in drinking water.
5. Next dry the prismatic tube using a lint-free cloth.
3 CEREC Bluecam
3.2 CEREC Bluecam with sapphire glass
23