3 CEREC Bluecam
3.2 CEREC Bluecam with sapphire glass
22
3.2.3

Removing the prismatic tube

A
Detent
1. Press the prismatic tube against the camera body.
2. Press detent A.
NOTE
Risk of damaging the front lens or prism.
➢ Push the prismatic tube straight toward the front, do not tilt it.
3. Pull off the prismatic tube.
3.2.4

Refitting the prismatic tube

NOTE
Do not use CEREC 2 / CEREC 3 prismatic tubes.
NOTE
Risk of damaging the front lens or prism.
➢ The prismatic tube must not touch the front lens.
➢ Push the prismatic tube straight toward the camera body; do not
tilt it.
➢ Carefully refit the prismatic tube until it locks in place.
3.2.5

Pre-cleaning of the prismatic tube

Cleaning process immediately after using the CEREC Bluecam: Wipe
the camera with the appropriate prismatic tube so that the superficial
impurities cannot harden and stick to the surface of the prismatic tube.
Remove any visible contaminations using cotton wool gauze, moistened
with tap water and a neutral soap, or lint-free cloth soaked with one of
the disinfectant agents specified in the "Cleaning agents and
disinfectants  [→ 3]" section. Wipe down the prismatic tube afterwards
again with the absorbent cotton gauze dipped in drinking water. Next
dry the prismatic tube using a lint-free cloth.
Instructions for CEREC Bluecam and CEREC Omnicam