I. Brightfield l\licroscopy
( 1) Push the mirror block changeover lever to the
click stop to put the brightfield mirror block (with
B.F. indication) in the optical path.
CONFIRM that the brightfield mirror block (B.F.)
is set at the right side of the mirror block port
(when seen from the front of the microscope).
(2) If the polarizer, the analyzer and the first order
red compensator are in the optical path, remove
them from the optical path.
*
During nonnal observation, Nomarski prisms SIA,
S 1 B, S 1 C or S 1 E can be left in the optical path. But
remove them when photomicrography is perfonned or
when a sample with much finer details is observed.
(3) Turn on the power switch of the power supply
unit. Rotate its lamp intensity control dial to
CTRL. Set the lamp intensity control dial of the
microscope base at about 9V.
(4) Push the anti-glare ND8 filter insertion/removal
lever to put this filter in the optical path.
I
s:
n
:ll
0
C/l
(")
.,
0
:ll
0
(")
m
0
C
:ll
m
24