® Eyepiece Tube
Sufficiently loosen the eyepiece tube clamp
screw at the top of the microscope with the
hexagonal screwdriver. Slightly tilt the eyepiece
tube to push its dovetail into the corresponding
mount. Tighten the clampscrew to fix the

eyepiece tube.

00D
Make sure that the serial number of the
eyepiece tube is identical to that of the
microscope base. The serial number of the
eyepiece tube is printed on the plaque on the
back of the microscope base.
® Eyepiece
Insert an eyepiece with a built-in mask (with
EPM indication) into the right eyepiece sleeve.
Insert an eyepiece with no built-in mask into the
left eyepiece sleeve. When inserting each
eyepiece, align its three grooves with the
corresponding protrusions of the eyepiece
sleeve. After inserting the eyepiece with a built­
in mask, confirm that this mask is not slanted.
7
, , , ,
I/
I/
I/
/'
,j
*
Rubber eyeguards for binocular
eyepieces (option)
Insert each eyeguard into the groove in the outer
circumference of the eyepiece.
*
Rotation clamp rings
The rotation clamp ring is attached to the
eyepiece sleeve to prevent the rotation of the
eyepiece. When removing one of the eyepieces,
do not hold the ring and forcibly pull it out
together with the eyepiece. The ring must be
detached when earlier-made eyepieces are
attached. Pay attention to the following when
attaching the ring again.
*
Attaching the rotation clamp ring
Hold the ring facing its ¢ 0.8 protrusion toward
you. There are two pawls on the ring and two
grooves on the sleeve. Attach the ring to the
sleeve to fit the pawls into the grooves.
A
t/J 0.8
protrusion
=-�
Pawl
v
-
Rotation clamp ring
Q)
Eyepiece sleeve