Functional Tests
Before Performing a Functional Test
Before Performing a Functional Test
1. Ensure that you have the proper test equipment.
2. Switch on the unit under test (UUT) and let it warm up (in accordance with warm-up requirements in
the instrument specifications).
3. Allow sufficient warm-up time for the required test equipment (refer to individual instrument
documentation for warm-up specifications).
4. Ensure that the analyzer's frequency reference is set to Internal:
a. Press the
Input/Output
b. If the
softkey does not show
Freq Ref In
5. Following instrument warm-up, perform the auto align routine:
Press
,
System
Alignments
NOTE
Functional test accuracy depends on the precision of the test equipment used. Ensure that
all of the test equipment is calibrated before running a functional test.
10
,
,
keys.
More
Freq Ref In
Internal
,
.
Align Now, All
, press the
softkey and select
Freq Ref In
.
Internal
Chapter 1