Frequency Response (Flatness), Preamp On

Test Limits
Frequency Range
100 kHz to 3.0 GHz
> 3.0 GHz to 7.5 GHz
The frequency response test, with preamplifier on, measures the signal analyzer's amplitude error as a
function of the tuned frequency. Measurements are made ranging from 50 MHz to the maximum
frequency range of the preamp. The signal source amplitude is measured with a power meter to eliminate
error due to source flatness. The measured value is normalized to 50 MHz.
Item
Adapter, Type N(m) to 3.5
mm(m)
Adapter,
3.5 mm(f) to 3.5 mm(f)
Cable, 3.5 mm, 1 meter
Cables, BNC 120 cm
Power Meter
Attenuator, 30 dB Fixed
Power Sensor
Power Splitter
Synthesized Sweeper
Limit Relative to 50 MHz
±2.0 dB
±3.0 dB
Critical Specifications
(for this test)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Frequency: 10 MHz
Compatible with power sensor
Frequency: 50 MHz
Accuracy: ±0.05 dB
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: –65 dBm to –55 dBm
Frequency Range: 50 MHz to 26.5 GHz
Tracking between ports: < 0.25 dB
Frequency Range: 50 MHz to 26.5 GHz
Recommended
Agilent Model
1250-1743
83059B
11500E
10503A
E4418B
11708A
8485D
11667B
PSG
35