Procedure

1. Zero and calibrate the power meter.
2. Configure equipment as shown in
analyzer input through the adapter.
CAUTION
To minimize stress on the test equipment connections, support the power sensor.
3. If the auto alignment for the analyzer has not been performed within the past 24 hours, press
,
Alignments
Align Now, All
4. Press
,
Mode
Spectrum Analyzer
5. Set up the synthesized sweeper by pressing:
, 50,
CW
MHz
, -4,
Power Level
dBm
(On)
RF
6. Set up the signal analyzer by pressing:
Input/Output, More, Freq Ref In
Channel, Center Freq, 50,
FREQ
, 2,
SPAN X Scale
MHz
,
AMPTD Y Scale
Attenuation, Mech Atten, 10, dB
,
Input/Output
RF Input
/Control, Sweep Setup,
Sweep
,
Meas Setup
Average/Hold Number
,
Trace/Detector
Trace Average
Single
7. Perform the following steps for each row listed in
a. Set the synthesized sweeper amplitude to the value listed in the Nominal Source Amplitude
column in
Table 5-1.
b. Set the Mech Atten as indicated in the Attenuation column in
c. Set the Span as listed in the Span column of
d. Record the source amplitude, as measured by the power meter, in the Power Meter Amplitude
column of
Table 5-1.
e. On the signal analyzer, press
f. Wait for the signal analyzer to finish averaging.
g. Press
.
Peak Search
h. Record the signal amplitude, as measured by the analyzer in the Measured Amplitude column of
Table 5-1.
Chapter 5
Figure
5-1., with the power splitter connected directly to the signal
to perform the auto alignment routine.
,
on the analyzer.
Mode Preset
, External
MHz
,
, select
RF Coupling
DC
,
Swp Time Rules
SA - Accuracy
, 20,
Enter
Table
Table 5-1.
.
Restart
Amplitude Accuracy at 50 MHz
5-1.:
Table 5-1.
Procedure
,
System