Procedure - Agilent Technologies X Series Manual

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Procedure

1. Configure the equipment as shown in
2. Press
,
Mode
Spectrum Analyzer
sweeper.
3. Set up the synthesized sweeper by pressing:
, 40,
Frequency
MHz
, –10,
Amplitude
dBm
4. Set up the signal analyzer by pressing:
Input/Output, More, Freq Ref In, External
,
FREQ Channel
Center Freq
,
SPAN X Scale
Span
5. On the analyzer, press
Peak Search
6. Adjust the synthesized sweeper amplitude for a signal analyzer display of –10 dBm ±0.1 dB.
7. On the analyzer, activate the marker delta function by pressing the
8. On the signal analyzer, press:
FREQ Channel, Center Freq,
,
Meas Setup
Average/Hold Number
Trace/Detector, Trace Average
Single
9. Press
. Enter the displayed value under the Measured Second Harmonic Distortion (dBc)
Peak Search
heading in
Table
4-1.
Table 4-1
Second Harmonic Distortion Results
Applied Frequency
Measured Second Harmonic Distortion
(MHz)
40
Chapter 4
Figure 4-1.
,
on the signal analyzer and Preset the synthesized
Mode Preset
, 40,
MHz
, 1,
MHz
.
80
, MHz
, 20,
Enter
(dBc)
Second Harmonic Distortion (SHD)
Procedure
and
keys.
Marker
Delta
Mixer Level
Specification
(dBm)
(dBc)
–20
–55
23