Displayed Average Noise Level (Danl) - Agilent Technologies X Series Manual

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

2

Displayed Average Noise Level (DANL)

Test Limits (with 0 dB input attenuation)
See
Table 2-1 on page 17
for values.
The Displayed Average Noise Level (DANL) of the signal analyzer is measured across a 10 kHz
frequency span at several center frequencies. The analyzer input is terminated into a 50Ω load. A test is
performed to assure the measurement is not performed in the presence of a residual response. The
measurement is then averaged, and the result is normalized to a 1 Hz bandwidth.
Critical Specifications
Recommended
Item
(for this test)
Agilent Model
Termination, 50Ω Type-N(m)
Frequency: DC to 18 GHz
909A Option 012
Figure 2-1.
DANL Test Setup
15

Advertisement

Table of Contents
loading

This manual is also suitable for:

N9000a cxa

Table of Contents