Procedure

1. Configure the equipment as shown in
2. Preset the Source and press
3. Set up the synthesized sweeper by pressing:
, 50,
Frequency
MHz
, +5,
Amplitude
dBm
, On
RF On/Off
4. Set up the signal analyzer by pressing:
,
Input/Output
More
FREQ Channel, Center Freq
SPAN X Scale, Span
AMPTD Y Scale, Ref Level,
Meas Setup, Average/Hold Number,
Trace/Detector, Trace Average
Peak Search
5. Set the external 10 dB step attenuator to 0 dB.
6. Adjust the amplitude on the signal source until the marker amplitude on the analyzer reads –15 dBm
±0.2 dB.
7. On the analyzer, press the
8. On the analyzer, activate the Marker Delta function by pressing
9. Perform the following steps for each attenuator setting listed in the table below:
a. Select the next External attenuator setting.
b. Press the
key to trigger a 10 sweep average.
Restart
c. Enter the delta marker value into
d. Check delta marker reading against the test limits.
Chapter 8
Figure 8-1.
,
Mode
Spectrum Analyzer
,
,
Freq Ref In
External
, 50,
MHz
1
,
, MHz
0
, dBm
10,
Enter
,
to trigger a 10 sweep average.
Single
Restart
Table
8-1.
,
on the analyzer.
Mode Preset
,
Peak Search
Marker Delta
Scale Fidelity
Procedure
.
41