The table below summarizes the test equipment needed to perform all of the functional tests. Alternate
equipment model numbers are given in case the recommended equipment is not available. If neither the
recommended nor the alternative test equipment are available, substitute equipment that meets or
exceeds the critical specifications listed.
Analyzer
Option
Item
Adapters
All
3.5 mm (f) to 3.5mm (f)
(connector saver for
source)
All
BNC (f) to SMA (m)
All
Type N (f) to Type N (f)
All
Type N (m) to 3.5 mm (m)
All
Type N (m) to 3.5 mm (f)
All
Type N (f) to 3.5 mm (f)
All
Type N (m) to BNC (f)
Attenuators
All
10 dB Step Attenuator
All
10 dB Fixed Attenuator (2
required)
Pxx
30 dB Fixed Attenuator
Cables
All
3.5 mm (1 meter)
Chapter 1
Critical Specifications
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 40 MHz
Frequency: 10 MHz to 18 GHz
VSWR: < 1.05:1
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 1 GHz
VSWR: < 1.05:1
Frequency: 50 MHz
Range: 0 to 70 dB
Frequency: 50 MHz
VSWR: < 1.2:1
Accuracy: < 0.05 dB @ 50 MHz
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Loss: < 2.0 dB
Functional Tests

Test Equipment

Alternate
Recommended
Agilent
Agilent Model
Model
1250-1749
83059B
1250-1200
1250-1472
1250-0777
1250-1743
1250-1744
1250-1745
1250-1476
8495A
8496A
8493C Option 010
8493A
Option 010
or
8493B
Option 010
11708A
11500E
8120-4921