Performance verification
62
2. Set the oscilloscope settings as follows:
Control
Vertical
Horizontal
Acquire
Trigger position
Trigger slope
Input impedance
Measure
3. Load the factory default preset. (See page 20, Load the factory preset.)
4. Use the oscilloscope to measure that the signal amplitude is within the range
of 900 mV to 1100 mV for the AES 1+2 channel.
5.
Record the measured value for AES 1+2 in the test record.
6. Use the oscilloscope to measure that the signal amplitude is within the range
of 900 mV to 1100 mV for the remaining AES channels as follows:
Change the BNC cable connection from the AES 1+2 connector to the
AES 3+4 connector and record the measured value for AES 3+4 in the
test record.
Change the BNC cable connection from the AES 3+4 connector to the
AES 5+6 connector and record the measured value for AES 5+6 in the
test record.
Change the BNC cable connection from the AES 5+6 connector to the
AES 7+8 connector and record the measured value for AES 7+8 in the
test record.
Change the BNC cable connection from the AES 7+8 connector to the
DARS connector and record the measured value for DARS in the test
record.
7. Change the BNC cable connection from the DARS connector to the AES
1+2 connector.
Setting
200 mV/div
100 ns/div
Sample
50%
Rising Edge
1 MΩ
Amplitude
SPG8000 Specifications and Performance Verification