Tektronix SPG8000 Technical Reference page 62

Master sync / clock reference generator
Hide thumbs Also See for SPG8000:
Table of Contents

Advertisement

Performance verification
48
3. Set the oscilloscope settings as follows:
Control
Vertical
Vertical offset
Horizontal
Horizontal position
Input
Trigger
Trigger level
Trig position
Trig slope
Acquire mode
4. Verify that the blanking level is within the range of +50 mV to –50 mV.
5. Record the measured value in the test record for Blanking Level.
6. Change the oscilloscope vertical scale to 10 mV/div.
7. Align the blanking level with the center graticule line on the oscilloscope.
8. Change the oscilloscope vertical offset to 300 mV.
9. Verify that the high level of the signal (sync amplitude plus) is within the
range of +0.6 div to –0.6 div to the center graticule (except for ringing of
the rising edge).
10. Record the measured value in the test record for Sync amplitude plus.
11. Change the oscilloscope vertical offset to –300 mV.
12. Verify that the low level of the signal (sync amplitude minus) is within the
range of +0.6 div to –0.6 div to the center graticule (except for ringing of
the falling edge).
13. Record the measured value in the test record for Sync amplitude minus.
14. Set the vertical offset to 0 V and change the vertical gain to 100 mV/div.
15. Repeat steps 4 through 13 for the Black 2 and Black 3 outputs, making sure to
move the cable to the connector of the output you are testing.
Setting
100 mV/div
0 V
500 ns/div
Center
CH1
CH1
–150 mV
50%
Rising Edge
Sample
SPG8000 Specifications and Performance Verification