Installing the analyzer modules and option module
5.2 Installation of the analyzer modules
5.2
Installation of the analyzer modules
5.2.1
Device expansion by a second analyzer module
If you install a second analyzer module, change the magnetic field frequency of the
OXYMAT 7:
If the device contains two OXYMAT 7 channels, set the magnetic field frequency of the last
analyzer module installed to 10 Hz.
28
Operating Instructions, 01/2013, A5E31930403-01
Wall-mounted device