Connecting
7.2 Connection to electricity supply
The mains supply line must comply with the valid regulations and conditions for the place of
installation and be provided with a protective conductor at the potential of the housing. The
cross-section of each wire has to be ≥ 1 mm
connected in the connector at the position marked.
● Ensure that a mains disconnection device is located in the direct vicinity of the device (for
● Ensure that the supply voltage present matches that which is specified on the nameplate
42
load rating, see nameplate).
of the device.
. The phased terminal conductors must be
2
Operating Instructions, 01/2013, A5E31930403-01
Wall-mounted device