Figure 10-1 Unlocking The Connector For Equipotential Bonding - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis, wall-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Replacing the sealing set
1. Isolate the device from power.
2. Open the door. To do this, unscrew the six screws.
3. If necessary, remove the installed analyzer module at the postion at which you want to
4. Remove all seals
5. Bond the new seals
Wall-mounted device
Operating Instructions, 01/2013, A5E31930403-01
CAUTION
Device damage
Before withdrawing the plug connector, loosen the lock. The contacts or the plug
connector could otherwise be damaged.
Figure 10-1
Unlocking the connector for equipotential bonding
replace the seal.
At the appropriate position, remove the blank plates and the plate with the cable glands
from the bottom of the housing.
using a suitable tool. Adhesive residues must be removed
completely.
previously removed plates using a suitable tool. Tightening torque: 0.8 Nm (analyzer
module and plates)
10.3 Replacing the door seal and sealing set
on again. If necessary, install the analyzer module. Tighten all
Maintenance and servicing
61