Installing the analyzer modules and option module

5.3 Installing the option module OM 2.1

5.3
Installing the option module OM 2.1
Preparing
Figure 5-5
1. Isolate the device from power.
2. Open the door. To do this, unscrew the six screws.
3. Remove the cover plate
4. Remove the connector from the power supply cable
34
Power supply cable
12-pin ribbon cable to AMs and option
module OM 2.1
Processing unit (PU)
26-pin ribbon cable to display
Cabling
the analyzer modules AM and to the option module OM2.1
cable to the display
the cables to the terminal block (TB)
. To do this, unscrew the three screws.
, from the 26-pin ribbon cable to analyzer module 2 (AM2)
⑦ ⑧
.
26-pin ribbon cable to AM2
Option module OM2.1
Cable between terminal block and
processing unit
Cable between AM2 and terminal block
, from the 12-pin ribbon cable to
, from the 26-pin ribbon
Operating Instructions, 01/2013, A5E31930403-01