Repair and replacement
October 2019
4. Remove and retain the M4 socket head cap screw (A), spring washer (B) and plain
5. Remove the O
6. Inspect the replacement O
7. Fit the replacement O
8. Connect the power cable by screwing it onto the mating connector on the O
9. Connect the BNC cable to the O
10. Close the door of the electrical compartment and restart the analyzer.
12.15.2
Replacing O
Procedure
1. Shut down the analyzer.
2. Open the door of the electrical compartment.
3. Disconnect the Bayonet Neill–Concelman (BNC) cable and the Power cable from the
148
Figure 12-12: Replace O
A. M4 socket head cap screw
B. M4 spring washer
C. M4 plain washer
D. O
detector sub-assembly
2
washer (C) used to secure the O
detector.
2
detector (D) in position using the M4 socket head cap screw
2
(A), spring washer (B) and plain washer (C) removed previously secure the O
detector to the base.
detector.
detector (type 2)
2
Refer to
Shutdown
procedure.
O
detector (D).
2
Refer to
Figure
12-13.
Detector (Type 1)
2
detector.
2
detector for damage.
2
detector.
2
Reference Manual