Reference Manual
00809-0100-4512
12
Repair and replacement
12.1
Repair policy
Repair and maintenance of the Rosemount
qualified personnel.
The analyzer has a number of parts that can be repaired and/or replaced with spare parts
supplied by Emerson.
In all cases the repair must be by direct replacement of the faulty item with a known
serviceable item purchased from Emerson.
NOTICE
Some faults can only be repaired by Emerson. Where an item is unserviceable, and no
replacement procedure is given in this manual, then the fault must be repaired by
Emerson.
12.2
Opening the electrical compartment door
To gain access to the electrical compartment and the components mounted on the inside
of the door, open the door on the front of the Rosemount CT5100(Ex) as follows:
Procedure
Using an 8 mm Allen key, loosen the compression screw (B) and cam latch screws (C) to
unlock the door (A), then open the door.
See
(Figure
Reference Manual
12-1).
Repair and replacement
™
CT5100(Ex) must only be performed by
October 2019
127