Emerson Rosemount CT5100 Reference Manual page 61

Continuous gas analyzer
Hide thumbs Also See for Rosemount CT5100:
Table of Contents

Advertisement

Reference Manual
00809-0100-4512
Table 5-3: Fuse Requirements
Fuse
2
3
4
5.4.3
Connecting the sample supply and return line
Sample gas supply and sample return connections are ¼-in. (6 mm) "Hamlet"
compression tube fittings. To avoid the risk of gas leaks, confirm that these connections
are made correctly and tightly. Both the sample gas supply pipe and the sample return
pipe should be thermal insulated.
The maximum sample gas supply pressure is 2 BarG.
5.4.4
Connecting the signal cables
The signal cables are connected to the system through an entry point outlet C, as shown in
Figure
Customer supplied conduit and cables are to be terminated in the entry point provided in
accordance with local electrical codes.
Reference Manual
Function
Neutral line 110/230
Vac (following mains
filter)
Analyzer 12 Vdc supply
rail
Analyzer 24 Vdc supply
rail
3-4. All signal cables are to be minimum 20 AWG tri-rated switchgear cable.
(continued)
Rating
3.15 A, 250 V, fast
acting ceramic
5 A, 250 V, fast acting
ceramic
3.15 A, 250 V, fast
acting ceramic
Install
October 2019
Schurter part number
0001.1009
0001.1011
0001.9011
61