6 Maintenance

6.4 Replacing the main fuse

82
Replacing the main fuse
6.4
DANGER
Potentially lethal shock hazard
People can be injured or electrical components of the unit destroyed.
➢ Switch off the unit prior to beginning work.
➢ Pull out the power cable.
NOTE
Fuse type
Use only fuses of the same type in the fuse module!
A
Main fuses
B
Fuse module
Fuses:
ü The power plug must be disconnected.
1. Unlatch the fuse module with a screwdriver and pull the module out.
2. Replace the defective fuses.
3. Reinsert the fuse module until it locks in place.
Operating Instructions CEREC Primescan AC, Primescan AC
C
Window
T8A H 250V
Dentsply Sirona
Order No. 62 33 188
66 54 573 D3696
D3696.201.01.05.02    09.2019