GRAPHTEC GL900 Quick Start Manual page 17

Midi logger
Hide thumbs Also See for GL900:
Table of Contents

Advertisement

16
Trigger Functions
Specify the condition as "Start data capture when the CH1 level exceeds 0.3V".
(1) Press the MENU key to open the "TRIG" menu and then set the Start Source
setting to "Level".
(2) Set the Combination to "Edge OR".
(3) Set the CH1 mode to "H".
(4) Open the Level Value Settings menu and set to "0.3V" with the cursor, and then
press the ENTER key.
(5) Align the cursor to the
ENTER key.
This completes the trigger settings.
Points to
Level detects whether or not it is
Remember !
above or below the set level.
Edge detects when it passes through
the set level.
button in the level menu settings and press the

Advertisement