4. Press the MENU key and open the "DATA" menu.
5. Set the sampling interval to "10ms".
Move the cursor to "Sampling" and then select "10ms".
6. Set the Data Capture Destination to "Internal RAM".
Move the cursor to "Data Capture Destination" and then select "Internal RAM".
Points to
Internal RAM, Internal flash memory and USB memory can be selected for
Remember !
Data Capture Destination.
Data can be recorded in the internal RAM at the sampling interval of 10μs or
more, but will be erased when the power is turned off or next capture is
performed. Internal flash memory and USB memory can record at the sampling
interval of 1ms or more for a long period of time (until the medium is full or one
file reaches 2GB).
7. Set the number of capture points to 10000.
Move the cursor to "Number of Capture Points" and then select "10000".
8. "Internal Flash Memory".
Move the cursor to "Auto Save" and then select "On".
Points to
Remember !
When data capture starts, auto save starts saving data in the auto save
destination at the same time with the internal RAM. Auto save is
continued in the background even when internal RAM data capture is
finished.