Sharing The M9301A Synthesizer's Lo - Keysight Technologies M9391A PXIe Startup Manual

Signal analyzer and vector signal generator 1 mhz to 3 ghz or 6 ghz
Hide thumbs Also See for M9391A PXIe:
Table of Contents

Advertisement

8x8 Shared LO Cable Connections:
For additional information, see

Sharing the M9301A Synthesizer's LO

For Phase-Coherent multiple channel operation, M9381A-012 option is required for
each PXIe VSG and M9391A-012 option is required for each PXIe VSA. For a single
chassis, multiple channel system, you can share a single M9301A PXIe Synthesizer
with up to four PXIe VSG or PXIe VSA instruments. For a multiple chassis, multiple
channel system, you can use Port 1a of the M9301A PXIe Synthesizer routed
through a V2802A LO Distribution Network to connect up to eight PXIe VSG or PXIe
VSA instruments.
Before enabling shared synthesizer functionality, install the modules and cable as
recommended in the previous sections.
59
Keysight M9391A PXIe Signal Analyzer & M9381A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
– Primary 100 MHz connection (8121-2175, long SMB-SMB); chassis
sync; 10 MHz distribution
– Secondary 100 MHz and IF connections (8120-5091, short SMB-SMB);
trigger cables
– LO distribution connections; Semi-rigids (V2800-20006, V2800-20008,
V2800-20009 for VSGs and V2800-20010, V2800-200011, V2800-20007 for
VSAs)
– PCIe cables (Y1202A)
– RF loopback
– Unused M9301A modules
Sharing the M9301A Synthesizer's LO
.

Advertisement