Step 4 - Install The Module; Step 4 - Install The Modules; Module Installation Process Overview - Keysight Technologies M9391A PXIe Startup Manual

Signal analyzer and vector signal generator 1 mhz to 3 ghz or 6 ghz
Hide thumbs Also See for M9391A PXIe:
Table of Contents

Advertisement

Step 4 - Install the Module

Step 4 - Install the Modules

Module Installation Process Overview

Keysight M9391A PXIe Signal Analyzer & M9381A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
PXIe hardware does not support "hot-swap" (changing modules while power is
applied to the chassis) capabilities. Before installing or removing a module to
/from the chassis, power off the chassis to prevent damage to the module.
All component modules for an M9381A or M9391A instrument are factory
tested, aligned, calibrated and shipped as a "bundle". It is important that you
maintain the bundle when installing modules or returning an instrument for
repair.
These modules can be used in a chassis with PXIe or PXIe-H chassis slots.
The module can be installed in any PXIe or hybrid PXIe slot marked with a peripheral slot
compatibility image (solid black circle for PXIe, or solid black circle with the letter -H- for hybrid).
Make sure that the line cord is plugged in to establish earth ground and that
1.
the chassis power switch is
If the chassis has multiple fan speed settings, ensure that the fan switch is
2.
set to
AUTO
and the inhibit switch is set to
Position the chassis to provide ample space between the chassis fan intake
3.
and exhaust vents. Blockage by walls or obstructions affects the air flow
needed for cooling. (Refer to the chassis documentation for more information
about cooling).
Before inserting a module into the chassis, back the mounting screws out to
4.
ensure that there is no interference between the screws and the mounting
rails.
Off
.
DEF
.
28

Advertisement

Table of Contents
loading

This manual is also suitable for:

M9381a pxie

Table of Contents