1
Calibration Procedures
Table 1-8
Close-case adjustment inputs (continued)
Test function
Reference value
12.209 mV
45.119 mV
Temperature
0 °C
[c] Ensure the clamp meter is turned on and stabilized for at least 1 hour with the thermocouple connected between the clamp meter and the
calibrator output terminal.
Vsense
9 V, 60 Hz
(Hi.SE)
Vsense
27 V, 60 Hz
(Lo.SE)
Exiting the adjustment mode
1 Remove all the shorting plugs and connectors from the instrument.
2 Power off and on again.
40
U1191A
-
-
[c]
-
-
-
U1192A
U1193A
U1194A
-
-
-
-
-
-
✔
✔
✔
✔
Keysight U1190A Series Service Guide
Valid reference input
0.9 to 1.1 × Reference value
✔
✔
0.9 to 1.1 × Reference value
0.9 to 1.1 × Reference value
✔
✔
N/A
N/A
✔