8 When all calibrations have completed, press both
9 Verify the adjustments using the
Exiting the adjustment mode
1 Remove all the shorting plugs and connectors from the multimeter.
2 Record the new calibration count.
3 Press both
4 Cycle the multimeter's power. The multimeter will then be secured.
Keysight U1241C/U1242C Service Guide
together for more than 1 second to exit the calibration mode. Alternatively,
you can also power off and on the multimeter again. The multimeter will return
to the normal measurement mode and is secured.
Log
Hold
calibration mode.
"Performance Verification Tests"
View
and
together for more than 1 second to exit the
Shift
Esc
Calibration Procedures
Log
View
and
Shift
Hold
Esc
on page 25.
1
51