Test Considerations - Keysight U1190A Series Service Manual

Handheld clamp meter
Hide thumbs Also See for U1190A Series:
Table of Contents

Advertisement

1
Calibration Procedures

Test Considerations

For optimum performance, all procedures should comply with the following
recommendations:
– The performance verification test or adjustment should be performed under a
– Ensure that the calibration ambient temperature is stable and is between 18 °C
– Ensure that the ambient relative humidity is less than 80%.
– The instrument should be in the laboratory environment for at least 1 hour
– Allow a warm-up period of 3 minutes.
– Use shielded twisted-pair PTFE-insulated cables to reduce settling and noise
24
laboratory condition, where the ambient temperature can be controlled.
and 28 °C. Ideally the calibration should be performed at 23 °C ± 1 °C.
prior.
errors. Keep the input cables as short as possible.
Keysight U1190A Series Service Guide