1
Calibration Procedures
5 Turn the rotary switch to the next function according to the Test Function
6 Verify the adjustments using the
While in the adjustment mode, turn the rotary switch to the OFF position to exit
NOTE
the adjustment mode.
36
column shown in
Table
shown in the adjustment table.
1-8. Repeat
step 2
to
"Performance Verification Tests"
Keysight U1190A Series Service Guide
step 4
for each adjustment point
on page 31.