Keysight U1190A Series Service Manual page 26

Handheld clamp meter
Hide thumbs Also See for U1190A Series:
Table of Contents

Advertisement

1
Calibration Procedures
Figure 1-2
1 Connect a dual banana plug to the output voltage terminals of the calibrator.
Ensure that the Fluke 5520A calibrator output is disabled when connecting the
NOTE
dual banana plug.
2 Set the Fluke 5520A calibrator output to the values shown in
3 Hold the clamp meter so that the top of the clamp meter is vertically and
Table 1-2
Vsense test
Step Test function
Vsense
1
Turn the rotary
switch to the
Press
to switch to Lo.SE.
4 Adjustment or repair is required if the clamp meter fails any of the Vsense
26
Vsense test
page 26, and enable the Fluke 5520A calibrator output.
horizontally centered and in contact with the banana plug's Hi terminal.
Range
position.
Lo.SE
functional tests. See
Dual banana plug
Positive output terminal
Fluke 5520A output
Hi.SE
18 V, 60 Hz
24 V, 60 Hz
"Close-case adjustment"
Fluke 5520A
Calibrator
Vsense alert status
5 V, 60 Hz
8 V, 60 Hz
on page 39.
Keysight U1190A Series Service Guide
Table 1-2
on
No alert
Alert on
No alert
Alert on