6 Options
6.1
ECG display
When the "ECG Display" option is activated, the ECG signal obtained by the elec-
trodes is displayed on the screen (see
• The "ECG Display" option is only available for the FRED easy® semi-automatic.
• This option is automatically activated when the FRED easy® is switched to manual
mode.
• When the ECG signal is displayed, messages are only displayed on one single
line.
6.2

Metronome

When the "Metronome" is enabled, the device sets the rhythm for chest compres-
sions during CPR at 100 bpm.
The FREDCO® software program allows you to:
– set the metronome frequency to a value between 80 and 150 bpm (in steps of
5 bpm).
– the metronome can be set to signal to the user to give 2 rescue breaths (the
metronome is stopped for 8 seconds after 30 chest compressions).
Fig. 6.1 ECG
display).
PRESS GREEN BUTTON
Fig. 6.1ECG display