Schiller AT-2 User Manual

Schiller AT-2 User Manual

12 channel ecg unit
Hide thumbs Also See for AT-2:

Advertisement

Quick Links

AT-2 User Guide
AT-2
AT-2
AT-2
12 Channel ECG Unit
AT-2
User Guide
1

Advertisement

Table of Contents
loading

Summary of Contents for Schiller AT-2

  • Page 1 AT-2 User Guide AT-2 AT-2 AT-2 12 Channel ECG Unit AT-2 User Guide...
  • Page 2 AT-2 User Guide CARDIOVIT AT-2 User Guide Art. No. 2.510 512 Rev b March 2004 SCHILLER AMERICA Inc., 11300 NW 41 Street, Miami, FL 33178 Telephone: +1 786 845 06 20 sales@schilleramerica.com www.schilleramerica.com...
  • Page 3: Intended Use

    The SCHILLER Cardiovit AT-2 is a 12-channel, ECG device used for the recording, analysis and evaluation of ECG Recordings. Recordings made with the AT-2 can be used as a diagnostic aid for heart function and heart conditions. The unit is designed for indoor use and can be used for all patients of both sexes, all races, and all ages.
  • Page 4: Table Of Contents

    AT-2 User Guide AT-2 User Guide Contents Intended Use ............................3 Physician`s Responsibility ........................3 Terms of Warranty ..........................7 Disposal Instructions and Battery Care ....................8 EMC Complicity ............................. 8 Power Supply ............................9 Changing a Mains Fuse ........................9 Safety Notices .............................
  • Page 5 AT-2 User Guide Keyboard ..............................18 Patient Cable Connections ......................... 20 Electrodes and Neutral Electrodes Identification and Color Code ............22 Electrode Placement - Standard Configuration ................... 23 Operating Overview ............................ 24 Automatic Mode ..........................24 Manual Mode ............................26 Changing the Lead Group ........................
  • Page 6 AT-2 User Guide Measurements and Markings (C version only) ..................44 Interpretation (C version only) ......................45 Interpretation Settings (C version only) ....................46 Selecting Rhythm Leads ........................47 Care & Maintenance ............................ 49 Self-test ............................... 49 12 Monthly Check ..........................49 Cleaning the Casing ..........................
  • Page 7: Terms Of Warranty

    Terms of Warranty The SCHILLER Cardiovit AT-2 is warranted against defects in material and manufacture for the duration of one year (as from date of purchase). Excluded from this guarantee is damage caused by an accident or as a result of improper handling. The warranty entitles free replacement of the defective part.
  • Page 8: Disposal Instructions And Battery Care

    Only dispose of the battery in official recycling centres or municipally approved areas. Alternatively, used batteries can be returned to SCHILLER AG for disposal. Units no longer required can be returned to SCHILLER AG for disposal. Alternatively dispose of the unit in municipally approved recycling centres.
  • Page 9: Power Supply

    AT-2 User Guide Power Supply The mains connection is on the rear of the unit. The power supply voltage is set by the factory for 100-115V(nom. 110V) working. The setting is indicated by the indented metal strip on the fuse panel. Contact your dealer if the voltage needs to be changed.
  • Page 10: Safety Notices

    This product is not designed for outdoor use. Surface temperature of applied parts must not exceed 41 Use only accessories and other parts recommended or supplied by SCHILLER AG. Use of other than recommended or supplied parts may result in injury, inaccurate information and/or damage to the unit.
  • Page 11: Safety Notices

    The device must only be operated using battery power if the earth connection is suspect or if the mains lead is damaged or suspected of being damaged. The AT-2 complies with emc regulations for medical products which affords protection against emissions and electrical interference. However special care must be exercised when the unit is used with high frequency equipment.
  • Page 12: Symbols And Conventions Used In This User Guide

    Ì CF Symbol. Unit is classified safe for internal and external use. The paddles at the side indicate that the unit is defibrillation protected when the original SCHILLER patient cable is used. Potential Equalisation Point. The unit /component can be recycled.
  • Page 13: Introduction

    AT-2 User Guide Introduction The SCHILLER Cardiovit AT-2 is an ECG recorder with all (12) ECG signals simultaneously processed to provide instant ECG recordings. Two automatic recording modes can be individually preset to enable one button ECG recording of preferred print formats.
  • Page 14 AT-2 User Guide...
  • Page 15: Modes Of Operation

    AT-2 User Guide Modes of Operation Automatic Mode Automatic Mode provides a printout giving 10 seconds of ECG recording of all 12 leads with a choice of 2 different formats. The following can be programmed freely for each of the 2 formats before recording: °...
  • Page 16: Location & Power

    AT-2 User Guide Location & Power Location Do not keep or operate the apparatus in a wet, moist, or dusty environment. Also, avoid exposure to direct sunlight or heat from other sources. Do not allow the unit to come into contact with acidic vapours or liquids, as such contact may cause irreparable damage.
  • Page 17: Switching On And Off

    Potential Equalisation If the AT-2 is used in conjunction with other patient connected equipment, we recommend that the potential equalisation stud on the rear of the unit is connected to the hospital/ building common ground with the yellow/green ground cable...
  • Page 18: Keyboard

    AT-2 User Guide Keyboard...
  • Page 19 PAPER Paper Jam or paper tray empty - replace paper FILTER Myogram filter applied. The numbers by the side of these keys are used when programming AT-2 system settings and auto mode formats etc. System settings are detailed later in this guide.
  • Page 20: Patient Cable Connections

    The Cardiovit AT-2 is CF rated. The patient connection is fully isolated and defibrillation protected. Protection against defibrillation voltage is however only ensured, if the original SCHILLER patient cable (Part-no.2.400071) is used. Make sure that during ECG recording neither the patient nor the conducting parts of the patient connection or the electrodes (including the neutral electrode) come into contact with other persons or conducting objects (even if these are earthed).
  • Page 21 AT-2 User Guide Patient Cable Connections The quality of the ECG is dependent on the preparation and the resistance between the skin and the electrode. To ensure a good quality ECG and minimise the skin/electrode resistance, remember the following points: Ensure that the patient is warm and relaxed.
  • Page 22: Electrodes And Neutral Electrodes Identification And Color Code

    AT-2 User Guide Electrodes and Neutral Electrodes Identification and Color Code The electrode placements shown in this handbook are labelled with the colors according to Code 2 requirements. The equivalent Code 1 colors are given below. CODE 1 (Usually European)
  • Page 23: Electrode Placement - Standard Configuration

    AT-2 User Guide Electrode Placement - Standard Configuration...
  • Page 24: Operating Overview

    AT-2 User Guide Operating Overview Automatic Mode In automatic mode, a full 12-lead ECG is printed in one of two predefined formats with a sensitivity of 10 mm/mV. These two formats are selected by the user to suit his specific needs and requirements.
  • Page 25 AT-2 User Guide Operating Overview The printout gives the following: ° ECG recording of all leads in either Standard or Cabrera format according to selection ° Sensitivity ° Heart Rate ° Speed ° Filter Settings ° Time and Date °...
  • Page 26: Manual Mode

    AT-2 User Guide Operating Overview Manual Mode Manual mode provides a direct printout of the real-time ECG with full control of parameter selection. To start the manual recording of a real-time ECG, press the MANUAL Printout key To stop the manual recording (printout) press the STOP key The printout provides you with the following: °...
  • Page 27: Changing The Lead Group

    AT-2 User Guide Operating Overview Changing the Lead Group The following can be freely chosen during the recording: Lead Group by means of the LEAD FORWARD and LEAD BACKWARD key The following lead groups are selectable: ° I, II, III...
  • Page 28: Chart Speed

    AT-2 User Guide Operating Overview Chart Speed Select speed 5, 25 or 50mm/s by means of the SPEED keys: Sensitivity Select 5, 10 or 20 mm/mV by means of the SENSITIVITY keys: Myogram Filter Switch the filter ON or OFF with the FILTER key...
  • Page 29: Settings

    AT-2 User Guide Settings Each parameter is set by means of a code. This code comprises a combination of keys starting with the ALT key followed by two or three numbers. The setting is confirmed with the STOP key. As soon as the ALT key is pressed, the keyboard is dedicated to the programming function.
  • Page 30 AT-2 User Guide Settings The defined formats and settings that are set for your unit can be checked as follows: Setup Printout Entry Key Sequence Result Printout of programmed Settings A printout of the defined settings will be produced and gives the following information, depending on the installed software:...
  • Page 31: Default Settings

    AT-2 User Guide Settings Printout of signals Sequential or Simultaneous Baseline Filter 0.05, 0.15 or 0.30 Hz Mains Filter 50, 60 Hz or OFF (-) Myogram Filter 25 or 35 Hz, ON (+) or OFF (-) Interpretation settings: N/A:+/- ‘normal/abnormal’ is written (+) or suppressed (-) U:+/- ‘unconfirmed report’...
  • Page 32 AT-2 User Guide Settings SETTINGS STANDARD WITH INTERPRETATION LANGUAGE AS SET AS SET LEADS STANDARD (S) STANDARD (S) ECG : 25MM/S, SHORT (O) MECG: 2*6 (50MM/S + 1) AUTO FORMAT 1 ECG: 25MM/S, SHORT (O) MEASUREMENTS: SUPRESSED (-) INTERPRETATION: ENABLED (+)
  • Page 33: Language

    AT-2 User Guide Settings Language Set the unit language as follows: Language Selection Entry Key Sequence Language Confirm German English French Swedish Press American STOP key Italian Spanish Portugese Russian NOTE: The difference between `English` and `American` is as follows:...
  • Page 34: Filters

    AT-2 User Guide Settings Filters There are three different filters which can be set individually as follows: ° Baseline filter ° Mains filter ° Myogram filter The setting for each filter is given on the setup printout.
  • Page 35: Baseline

    AT-2 User Guide Settings Baseline Baseline Filter The baseline filter greatly reduces the baseline fluctuations without affecting the ECG signal. The purpose of the this filter is to keep the ECG-signals on the baseline of the printout. This filter is only effective in auto mode printout.
  • Page 36: Mains Filter

    AT-2 User Guide Settings Mains Filter The Mains filter is an adaptive digital interference filter designed to suppress AC interference without attenuating or distorting the ECG. Set the mains filter in accordance with the frequency of your local mains supply as follows:...
  • Page 37 AT-2 User Guide Settings Myogram Filter Entry Key Setting Confirm Sequence Myogram Filter 25 Hz Myogram Filter 35 Hz Myogram Filter active Press when the unit is first STOP switched on (marked on printout with +) Myogram Filter off when the unit is first...
  • Page 38: Defining Lead Sequence & Printout

    AT-2 User Guide Settings Defining Lead Sequence & Printout The required settings can be selected as follows: Sequences, Print & Auto-centering Entry Key Definition Confirm Sequence Standard Lead Sequence Cabrera Lead Sequence Press STOP Simultaneous Print Sequential Print Auto-centering ON...
  • Page 39 Auto-Centering ON All ECG traces are centred dynamically for optimal use of paper width. Auto-Centering OFF ECG traces are set to a fixed baseline position and may possibly overlap. The Standard and Cabrera lead groups available for the AT-2 are: Lead Groups Standard...
  • Page 40: Acoustic Qrs Indication

    AT-2 User Guide Settings Acoustic QRS Indication The acoustic QRS beep can be switched on or off at any time by pressing the QRS key Time / Date The required settings can be selected as follows: Setting the Time and Date...
  • Page 41: Automatic Mode (Ecg) Settings

    Automatic Mode (ECG) Settings Two separate Auto formats can be defined for the AT-2. When defining auto format 1 the key sequence ALT `1` precedes the setting. When defining auto format 2 the key sequence ALT `2` precedes the setting.
  • Page 42 AT-2 User Guide Automatic Mode (ECG) Settings The automatic mode formats are detailed on the following pages. The ECG format is set as follows: ECG Format Entry Key Sequence Printout Confirm 1page x 12 leads at 25mm/s One page with the first 8 leads...
  • Page 43: Average Cycles

    AT-2 User Guide Automatic Mode (ECG) Settings Average Cycles ( C version only) Average Cycles Entry Key Sequence Printout Confirm The printout format for the Average cycles are given on the No average lead table cycles are printed 4 x 3 (25 mm/s) + 2 rhythm leads (25mm/s).
  • Page 44: Measurements And Markings (C Version Only)

    AT-2 User Guide Automatic Mode (ECG) Settings Measurements and Markings (C version only) To define the measurements and markings proceed as follows: Measurements (Interpretation Option Only) Entry Key Sequence Printout Confirm Detailed table of measurement results omitted - however, the...
  • Page 45: Interpretation (C Version Only)

    Interpretation (Interpretation Option Only) Entry Key Sequence Printout Confirm Interpretation is Press omitted 1 or 2 STOP Interpretation is printed Full details of the interpretation option are given in the SCHILLER ECG Measurement and Interpretation booklet (art. No. 2.510 179).
  • Page 46: Interpretation Settings (C Version Only)

    AT-2 User Guide Automatic Mode (ECG) Settings Interpretation Settings (C version only) The interpretation settings enable the user to determine whether or not certain comments will be added to the interpretation statements on the ECG printout. Furthermore, the patient’s age can be defined (<30 or >30) and if low or high sensitivity should be applied.
  • Page 47: Selecting Rhythm Leads

    AT-2 User Guide Automatic Mode Settings Selecting Rhythm Leads The rhythm leads are printed out as defined. Two separate rhythm leads can be selected. The following formats can be set: Rhythm Leads (interpretation option only) Entry Key Setup Format Sequence...
  • Page 48 AT-2 User Guide Automatic Mode Settings The 2 rhythm leads are defined as follows: Extremity Leads Entry Key Sequence Lead Confirm Press 3 or 4 STOP Precordial Leads Entry Key Sequence Lead Confirm Press 3 or STOP...
  • Page 49: Care & Maintenance

    AT-2 User Guide Care & Maintenance Self-test Initiate a self-test of the AT-2 as follows: Self Test Entry Key Sequence Action Service Data Displayed A table giving information for the service staff is displayed. To obtain a printout press `P` when the table is displayed. Exit this screen by pressing the ENTER key.
  • Page 50: Cleaning The Casing

    The casing of the AT-2 can be cleaned with a soft damp cloth on the surface only. Where necessary a domestic non- caustic cleaner can be used for grease and finger marks.
  • Page 51: Cleaning The Patient Cable

    AT-2 User Guide Care & Maintenance Cleaning the Patient Cable WARNING Align the leads in such a way as to prevent anyone stumbling over them or any damage caused by the wheels of instrument trolleys. The patient cable should not be exposed to excessive mechanical stress. Whenever disconnecting the leads, hold the plugs and not the cables.
  • Page 52: Replacing The Recording Paper

    AT-2 User Guide Replacing the Recording Paper The recording paper must be replaced as soon as the end of the paper is indicated by a red stripe on the lower edge. After the indication first appears, there are about 8 pages left. However, we recommend that the paper be replaced immediately.
  • Page 53: Thermal Paper Handling

    Thermal Paper Handling The thermal paper used in the AT-2 requires slightly different handling to normal paper as it can react with chemicals and to heat. However, when the following points are remembered, the paper will give reliable results: The following points apply to both storage, and when archiving the results.
  • Page 54: Fault Diagnosis

    No? Check mains supply. Yes? Press the OFF key Wait a few seconds and switch on again. Check fuses - Call your local SCHILLER representative. QRS traces overlap Ensure that the automatic sensitivity reduction is not switched off. Reset signals to baseline - press the 1mV key Check electrode contact ‘Noisy’...
  • Page 55 Ensure that fresh SCHILLER paper is installed. Note that the thermal paper used for the AT-2 is heat and light sensitive. If is not stored in its original seal, stored in high temperatures or is simply old, print quality can deteriorate.
  • Page 56: Ordering Information

    Ordering Information Your local representative stocks all the disposables and accessories available for the AT-2. In case of difficulty or to obtain the address of your local dealer, please contact the head office. Our staff will be pleased to help process your order or to provide any details for all SCHILLER products.
  • Page 57: Technical Data

    AT-2 User Guide Technical Data Technical data subject to change without notice. Dimensions 400 x 330 x 100 mm Weight 5.0 kg ( 5.35 kg with full paper tray) Mains Supply 100 to 115 / 220 to 240 VAC, 50/60 Hz...
  • Page 58 AT-2 User Guide Technical Data ECG Storage: Circular input memory for 10 s, 12-lead ECG. Frequency Range of Digital Recorder: 0 to 150 Hz (IEC) 0 to 150 Hz (AHA) ECG Amplifier: Simultaneous, synchronous registration of all 9 active electrode signals (= 12 standard...
  • Page 59 AT-2 User Guide Technical Data Patient Input: Fully floating and isolated, defibrillation protected. Safety Standard: CF according to IEC and complying with the following RL 93/42/EEC EN 60601-1:1990 IEC 601-1 IEC 601-2-25:1993 pr EN 1441:1994 EMC: CISPR 11: 1985, EN 55011: 1992...
  • Page 60: Available Configurations

    Control Panel: Rubber keys Technical data subject to change without notice. Available Configurations The Cardiovit AT-2 is available in two versions: Standard Version: Unit with ECG recording and printout capabilities. Version C: Unit with additional ECG Interpretation program (including measurements).

Table of Contents