Trouble Shooting; Error Messages - Schiller FRED easy User Manual

Automated external defibrillator (aed)
Hide thumbs Also See for FRED easy:
Table of Contents

Advertisement

FRED easy
Error
INFO: XXX
PERFORM
CHEST COMPRESSIONS
Error message
TEST FAILURE
XXX
• xxx =
– ADC
– EEPROM
– RTC
– LCD
– OKI
– DSP
– SHOCK BUTTON
– CHARGE
DEFIBRILLATOR
– COMMUNICATION
DEFIBRILLATOR
– 5 WEEKS
CHARGE 150J
Error
INFO: XXX
• xxx =
– CPU DEFI
– CRC DEFI
– SAFETY DEFI
– REF VOLTAGE DEFI
– ADC DEFI
– DEFI DISCHARGE
– DEFI EPROM
– DEFI SHOCK KEY
User Guide
7.5

Trouble Shooting

• If it is not possible to get the device back into operating condition within a reason-
able period of time, continue cardiopulmonary resuscitation until the rescue service
arrives.
• If the device switches to Limited mode (CPR) due to an error, the alarm messages
may differ.
Forced shutdown procedure
• If the device cannot be switched off via normal OFF procedure (press and hold the
green button for approx. 3 seconds), remove the battery and inserting again.
7.5.1

Error messages

Possible cause
• Technical errors that may occur
during function tests:
• Technical problems that can oc-
cur during defibrillation
Maintenance
Trouble Shooting
Remedy
 Switch the unit off and then on again to con-
firm. If the message still appears, the device
must be repaired immediately.
 Switch the unit off and then on again to con-
firm. If the message still appears, the device
must be repaired immediately. Continue
cardiopulmonary resuscitation until the res-
cue service arrives.