3.2 Electrical Safety

General specifications
Microscope
For indoor use only.
Input:
Supply voltage:
Max. Power
Frequency:
Fuses:
Output:
Output voltage:
Max. Power:
Ambient temperature: 15-35°C
Relative humidity:
Overvoltage category: II
Pollution degree:
WARNING
Use only original power cables or alterna-
tive cables with a VDE / HAR logo, which at
least fulfill the requirement of 3x0.75 mm
10A/250V.
Caution!
Plugging or unplugging of cables is to take
place in the de-energized state!
100-240V AC
18VA
50/60Hz
5x20, T1.6A H 250 VAC
5V DC
10W
max 80% to 30°C
2
and
2
WARNING
By definition, the main circuit breaker of this
instrument is the connection between the
power cable and device port. The user must
ensure unobstructed access to the main cir-
cuit breaker at all times.
WARNING
The power plug may only be plugged into an
outlet equipped with a grounding contact.
Do not interfere with the grounding function
by using an extension cord without a ground
wire. Any interruption of the ground wire in-
side or outside of the device, or release of
the ground wire connection, can cause the
device to become hazardous. Intentional
ground interruption is not permitted!
Do not use this instrument near sources of
high electromagnetic radiation (for example,
unshielded, intentionally operated ultra-high
frequency sources), because these can dis-
rupt proper operation.
We recommend assessing the electromag-
netic environment before operation of this
instrument and then giving corresponding
instructions.
3. Safety Notes